ORIGINAL ARTICLE
Year : 2020  |  Volume : 9  |  Issue : 1  |  Page : 68-70

Efficacy of auto-PAP titration in obstructive sleep apnea: Single-center experience


1 Department of Respiratory Therapy, Amrita Institute of Medical Sciences, Kochi, Kerala, India
2 Department of Respiratory Therapy, Manipal College of Health Professions, Manipal, Udupi, Karnataka, India
3 Department of Pulmonology, VPS Lakeshore Hospital and Research Centre, Eranakulam, Kerala, India

Date of Submission17-Jul-2019
Date of Decision13-Nov-2019
Date of Acceptance14-Nov-2019
Date of Web Publication08-Jan-2020

Correspondence Address:
Mr. G Gokul Krishna
Department of Respiratory Therapy, Manipal College of Health Professions, Manipal, Udupi, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijrc.ijrc_35_19

Rights and Permissions
  Abstract 


Background: Obstructive sleep apnea (OSA) is a sleep-associated breathing disorder, which left untreated can cause severe morbidities or even mortality. The recommended treatment strategy for moderate-to-severe OSA is continuous positive airway pressure (CPAP). The optimal pressure required for OSA can be measured by either manual CPAP titration or auto-CPAP titration study. Objective: To assess the efficacy of auto-PAP titration for OSA patients. Methodology: This was a cross-sectional study conducted on Pulmonary Medicine Sleep Lab at Amrita Institute of Medical Sciences, Kochi. Fifty patients who have undergone full-night polysomnography and followed with auto-PAP titration are included in the study. The research was approved by the institutional ethics committee. Results: In the study cohort, 50% showed optimal titration, 40% good titration, 10% adequate titration and none in unacceptable category. The results showed that auto-PAP usage has good result in moderate-to-severe OSA. Conclusion: Unattended auto-PAP titration seems to be highly effective modality which can be considered as an alternative to attended CPAP titration, thus reducing labor intensiveness and cost.

Keywords: Auto-positive airway pressure, continuous positive airway pressure, OSA


How to cite this article:
Unnithan AS, Krishna G G, Lakshmanan P H. Efficacy of auto-PAP titration in obstructive sleep apnea: Single-center experience. Indian J Respir Care 2020;9:68-70

How to cite this URL:
Unnithan AS, Krishna G G, Lakshmanan P H. Efficacy of auto-PAP titration in obstructive sleep apnea: Single-center experience. Indian J Respir Care [serial online] 2020 [cited 2020 Jun 2];9:68-70. Available from: http://www.ijrconline.org/text.asp?2020/9/1/68/275380




  Introduction Top


Obstructive sleep apnea (OSA) is characterized by partial or full obstruction of the airway during sleep due to muscle relaxation causing soft-tissue collapse in the upper airway. Continuous positive airway pressure (CPAP) is the recommended treatment for OSA; it acts as a pneumatic splint in the airway and keeps it open.[1] CPAP level required can be assessed either by manual CPAP titration or auto-titration. In manual titration, an attending clinician monitors the sleep pattern and increases or decreases CPAP level manually. In auto-PAP titration, the machine automatically adjusts for varying pressure requirement of patients. During the course of auto-PAP titration, one attending clinician should monitor the patient and minimize the chances of disconnection. Auto-PAP devices adjust treatment pressure by feedback control according to patterns of sleep and pressure requirement. Flow signals are measured during the treatment.

According to the American Academy of Sleep Medicine (AASM) guidelines, efficacy of CPAP titration is classified as follows: Optimal titration: Reduces Respiratory Disturbance Index (RDI) <5 for at least a 15 min duration and should include supine rapid eye movement (REM: During this stage, the activity of the brain's neurons is quite similar to that during waking hours) sleep at the auto-titrated pressure that is not continually interrupted by spontaneous arousals or awakening. Good titration: Reduces RDI <10 or by 50% if the baseline RDI <15 and should include supine REM sleep that is not continually interrupted by spontaneous arousals or awakening. Adequate titration: Does not reduce RDI <10 but reduces the RDI by 75% from baseline, or one which the titration grading criteria for optimal or good are met with the exception that supine REM sleep did not occur at the auto-titrated pressure. Unacceptable titration: Titration is one that does not meet any one of the above grades.[2]


  Methodology Top


The present observational study was conducted over a period of 1 year (August 2016–August 2017) at the Pulmonary Medicine Department in Amrita Institute of Medical Science and Research Center, a multispecialty teaching and tertiary care hospital, Kochi, India. The study was approved by the Institutional ethical committee. Consecutive patients aged >18 years, who were diagnosed with moderate (apnea-hypopnea index (AHI) 15-29) to severe OSA (AHI ≥30) with polysomnography (PSG) and who underwent auto-PAP titration were included in the study. Patients with overlapping conditions such as chronic lung disease or who were acutely ill (heart failure), mild OSA, already on noninvasive ventilation, and were not willing to participate from the study were excluded. Auto-PAP level ranged between 4 cmH2O and 20 cmH2O. The pressure was automatically titrated by the auto-PAP machine. Philips Respironics Alice PDx and Resmed AirSense 10 auto-CPAP devices were used for the study.

Outcome measures

The primary outcome assessed was the efficacy of auto-PAP titration in moderate-to-severe OSA. Secondarily we also assessed association between body mass index (BMI), age distribution, neck circumference, and leak with titration efficacy groups.


  Results Top


[Table 1] represents the demographic details of the patients. 50 patients were recruited in the study [Figure 1]. Of these, 45 were males. 73% of the study groups were obese (BMI ≥30). Based on age, the patients were stratified into three groups, of which 58% of the study population with OSA had age between 40 and 59 years, respectively. PSG study showed 28% of study population had moderate OSA and 72% had severe OSA. [Table 2] represents the comparison of OSA severity with other measured indices. The mean respiratory distress index (RDI) was 25.5 vs. 48.8 in moderate and severe OSA group. Higher percentage of both moderate and severe OSA patients had more than 4 Epworth score for daily symptoms. Average snore index in patients with moderate and severe OSA was 214 and 354, respectively. The primary outcome variable (efficacy of CPAP) assessed with AASM guideline showed 25 (50%) patients had optimal titration, 20 had good, and 5 had adequate titration. No one came under the category of unacceptable titration [Table 3]. Secondary assessments such as comparison of titration group versus the age distribution, BMI, neck circumference, and leak were measured and are reported in [Table 4] and [Table 5].
Table 1: Demographic data

Click here to view
Figure 1: Schematic representation of study design

Click here to view
Table 2: Obstructive sleep apnea severity versus other measured data

Click here to view
Table 3: Auto-positive airway pressure titration efficacy

Click here to view
Table 4: Age distribution versus titration efficacy

Click here to view
Table 5: Mean body mass index, neck circumference and leak in titration group

Click here to view



  Discussion Top


OSA is a common disease condition with high prevalence in developed and developing countries. Epidemiological studies from USA showed almost 20 percent of adults have at least mild OSA and 7 percent of them have moderate-to-severe OSA with male predominance.[3] Furthermore, data from Indian home sleep studies showed 7.5% of urban middle age men have OSA. Obesity and aging were the risk factors for OSA. Similar to these studies, the present study also observed the incidence of OSA was higher in males, aged between 40 and 59 years. One of the earliest studies on treating OSA with CPAP was done by Sullivan et al. in South Asian population.[4] In their study, they observed significant difference in AHI with application of CPAP. Following this, the benefits of CPAP in OSA have been well studied and documented to be an effective strategy.[5],[6],[7],[8],[9] Later, auto-PAP titration was introduced and has been studied extensively. Ortho et al. conducted a cohort study comparing the constant pressure with automatic pressure and found auto-PAP was as effective as constant CPAP in OSA.[10] Ficker et al. conducted a study on auto-PAP versus constant CPAP and showed that auto-PAP was equally effective as manual CPAP.[11] Another interesting finding was that the level of CPAP was higher in auto-PAP group compared to the constant CPAP. The primary objective of the present study was to assess the efficacy of auto-PAP. We found that 50% were optimally titrated, 40% showed good titration, 10% titrated adequately, hardly anyone in unacceptable efficacy category.

The secondary objective was to compare the association between BMI, age, and neck circumference and leak statistics with titration efficacy. We observed that the mean BMI, neck circumference, and leak were similar compared to the titration group. Ress et al.[12] and Mayer et al.[13] showed that most of the baseline parameters have significant correlation with titration efficacy.

Major limitation of our study was lack of a comparative cohort, but study findings support the use of auto-PAP titration to reduce AHI. The scarcity of manpower resource is a major limitation of delivering quality care to patients and therefore in resource-limited settings, auto-CPAP titration can be recommended over manual CPAP titration.


  Conclusion Top


Unattended auto-PAP titration seems to be a highly effective modality that can be considered as an alternative to attended CPAP titration. Thus reduces labor intensiveness and cost.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Sforza E, Krieger J, Bacon W, Petiau C, Zamagni M, Boudewijns A. Determinants of effective continuous positive airway pressure in obstructive sleep apnea. Role of respiratory effort. Am J Respir Crit Care Med 1995;151:1852-6.  Back to cited text no. 1
    
2.
Kushida CA, Chediak A, Richard B, Brown LK, Gozal D, Iber C, et al. Clinical guidelines for the manual titration of positive airway pressure in patients with obstructive sleep apnea positive airway pressure titration. J Clin Sleep Med 2008;4:2.  Back to cited text no. 2
    
3.
Young T, Peppard PE, Gottlieb DJ. Epidemiology of obstructive sleep apnea: A population health perspective. Am J Respir Crit Care Med 2002;165:1217-39.  Back to cited text no. 3
    
4.
Sullivan CE, Issa FG, Berthon-Jones M, Eves L. Reversal of obstructive sleep apnoea by continuous positive airway pressure applied through the nares. Lancet 1981;1:862-5.  Back to cited text no. 4
    
5.
Loube DI, Gay PC, Strohl KP, Pack AI, White DP, Collop NA. Indications for positive airway pressure treatment of adult obstructive sleep apnea patients: A consensus statement. Chest 1999;115:863-6.  Back to cited text no. 5
    
6.
Indications and standards for use of nasal continuous positive airway pressure (CPAP) in sleep apnea syndromes. American Thoracic Society. Official statement adopted March 1944. Am J Respir Crit Care Med 1994;150:1738-45.  Back to cited text no. 6
    
7.
Jenkinson C, Davies RJ, Mullins R, Stradling JR. Comparison of therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: A randomised prospective parallel trial. Lancet 1999;353:2100-5.  Back to cited text no. 7
    
8.
Engleman HM, Kingshott RN, Wraith PK, Mackay TW, Deary IJ, Douglas NJ. Randomized placebo-controlled crossover trial of continuous positive airway pressure for mild sleep Apnea/Hypopnea syndrome. Am J Respir Crit Care Med 1999;159:461-7.  Back to cited text no. 8
    
9.
Oksenberg A, Silverberg DS, Arons E, Radwan H. The sleep supine position has a major effect on optimal nasal continuous positive airway pressure: Relationship with rapid eye movements and non-rapid eye movements sleep, body mass index, respiratory disturbance index, and age. Chest 1999;116:1000-6.  Back to cited text no. 9
    
10.
d'Ortho MP, Grillier-Lanoir V, Levy P, Goldenberg F, Corriger E, Harf A, et al. Constant vs. automatic continuous positive airway pressure therapy: Home evaluation. Chest 2000;118:1010-7.  Back to cited text no. 10
    
11.
Ficker JH, Wiest GH, Lehnert G, Wiest B, Hahn EG. Evaluation of an auto-CPAP device for treatment of obstructive sleep apnoea. Thorax 1998;53:643-8.  Back to cited text no. 11
    
12.
Rees K, Wraith PK, Berthon-Jones M, Douglas NJ. Detection of apnoeas, hypopnoeas and arousals by the AutoSet in the sleep apnoea/hypopnoea syndrome. Eur Respir J 1998;12:764-9.  Back to cited text no. 12
    
13.
Mayer P, Meurice J, Philip-Joet F, Cornette A, Rakotonanahary D, Meslier N, et al. Simultaneous laboratory-based comparison of ResMed Autoset (TM) with polysomnography in the diagnosis of sleep apnoea/hypopnoea syndrome. Eur Respir J 1998;12:770-5.  Back to cited text no. 13
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methodology
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed299    
    Printed22    
    Emailed0    
    PDF Downloaded63    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]